Pulse type turbine flowmeter

6液体涡轮流量计(脉冲输出).png
6气体涡轮流量计(脉冲输出).png
  • 6液体涡轮流量计(脉冲输出).png
  • 6气体涡轮流量计(脉冲输出).png
Pulse type turbine flowmeter is adopted advanced ultra low power single chip microcomputer technolog…

Description:

Products overview:
ZW - LW seriesPulse type turbine flowmeterIs a kind of precision flow measuring, has simple structure, light, high precision, good repeatability, responsive, use of installation and maintenance is convenient wait for a characteristic of a new generation of turbine flowmeter, and the corresponding supporting flow integrating instrument can be used to measure the instantaneous flow and cumulative flow of liquid.Can also control quantitatively, excessive alarm, etc., is an ideal instrument to flow measurement and energy saving.Widely used in petroleum, chemical industry, metallurgy, scientific research in the areas of measurement and control system.Equipped with health joints threaded connection type liquid turbine flowmeter sensor can be used in the pharmaceutical industry.

Second, the working principle:
ZW - LW seriesThe principle of pulse type turbine flowmeter is a fluid flowing through the sensor shell, as a result of the impeller blade and flow to a certain point of view, the impact of fluid blade with rotational torque, to overcome the friction torque and fluid resistance, blade rotation speed stability after the moment balance, under certain conditions, the speed is proportional to the velocity of flow, because there are magnetic conductivity blade, it is in the signal detector (composed of * magnetic steel and coil) magnetic field, rotating blade cutting lines, periodically changing coil magnetic flux, so that the coil ends induction electrical pulse signal, the signal after amplifier amplification plastic, form a certain amplitude of continuous rectangular pulse wave, can be spread far to display instrument, show the instantaneous flow and cumulative amount of fluid.Within the scope of a certain flow, pulse frequency f and flows through the sensor of fluid is proportional to the instantaneous flow rate Q, flow equation is: Q = 3600 x f/k;Type:
F - pulse frequency (Hz);
K - the instrument coefficient of sensor 1 / m3, given by the check list.If [1 / L] as the unit
Q - fluid transient flow (working conditions) (m3 / h),
3600 - conversion factor.
Coefficient of each sensor of the instrument in the calibration certificate will be completed by the manufacturer, set into the display instrument of form a complete set of k value, can show the instantaneous flow and cumulative total.
Applications:
ZW - LW seriesPulse type turbine flow meter with high precision, good repeatability, simple structure, less moving parts, high pressure resistance, wide measuring range, small volume, light weight, small pressure loss, easy maintenance, etc, used for measuring low viscosity the volume of a gas in a closed pipe flow and total amount.In petroleum, chemical industry, metallurgy, urban gas network industries has wide use value.
Four, the main features:
1, wide measuring range, the lower velocity of less than 0.5 m/s, small pressure loss, the impeller impact resistant ability.
2, has the high resistance to electromagnetic interference and vibration resistance, self-lubricating bearing adopts full sealing isolation protection, performance reliable working life is long.
3, USES the advanced ultra low power single chip microcomputer technology, the whole machine function is strong, low power consumption, superior performance.With nonlinear compensation function of intelligent traffic monitor precision.Accuracy is better than that of plus or minus 0.02%.
4, meter coefficient can be set up by the key online and can be displayed on the LCD screen, LCD screen clear intuitive, high reliability.
5, block the EEPROM for cumulative flow, instrument coefficient of power-fail protection.To protect more than 10 years.
6, can be in the measured gas pressure compensation under stable state of stress.
Five,Technical parameters: 
1, the nominal diameter: pipeline: DN4 DN200 ~;Insert: DN100 ~ DN2000; 
2, precision grade: pipeline: plus or minus 0.5, plus or minus 1.0 on the Richter scale;Plug-in: 1.5 mm, 2.5 mm; 
3, environmental temperature: - 20 ℃ to 50 ℃; 
4, medium temperature: liquid measurement: - 20 ℃ ~ 120 ℃;Measuring gas: - 20 ℃ ~ 80 ℃; 
5, the atmospheric pressure: 86 kpa to 106 kpa. 
6, nominal pressure: 1.6 Mpa, 2.5 Mpa, 6.4 Mpa, 25 Mpa; 
7, explosion-proof level: ExdIIBT4; 
8, connection mode: threaded connection, flange clamp outfit, flange connection, plug, etc; 
9, straight pipe requirements: gas: a straight pipe upstream should be 10 or more dn, downstream straight pipe should be 5 or more dn. 
Liquid: straight pipe upstream should be 20 or more dn, downstream straight pipe should be 5 or more dn. 
Plug-in: straight pipe upstream should be 20 or more DS, downstream straight pipe should be 7 or higher DS (DS for pipeline measured diameter); 
10, display mode: (1) far eastone show: pulse output, the output current (match display instrument); 
(2) the scene showed that eight LCD display cumulative flow, unit (m3); 
Four transient flow, LCD display unit (m3 / h), battery, frequency, flow velocity; 
(3) the temperature and pressure compensation type: 
A, according to standard criteria for instantaneous flow and cumulative flow; 
B, show the current pressure, temperature, battery voltage; 
11, output functions: 
(1) pulse output, p - p values determined by power supply; 
Two wire (2) 4 ~ 20 ma current output; 
(3) pulse output per unit volume and the original pulse sensor output; 
(4) with RS485 communication interface; 
12, power supply: 
(1)DC5~24V; 
(2) the standard 3 v lithium battery installed on the instrument internal can use eight consecutive years; 
(3) temperature pressure compensation type 3 v lithium battery installed within the instrument can be continuous use of more than 4 years; 
13, transmission distance, sensors to display the distance up to 500 m.

Six, wiring and grounding considerations: 
Wiring matters needing attention 
1, using a full cable as soon as possible. 
2, should as far as possible away from electrical noise (such as power transformer, electric motor and the power cord), avoid parallel to the power the power cord wiring. 
3, Suggestions in the end of the thick wire adopts the tin welding clamping lugs. 
4, for the waterproof and mechanical damage, the cable into the metal tube inside, but may not have high power transmission cables, within the same catheter (a transmission cable transmission of zui power is greater than the flowmeter signal transmission cables zui small power 10 times, both cannot be installed on the same catheter). 
5, the existing strong magnetic field outside the area, should make the axis of the detection device with the outside world vertical or the direction of the magnetic flux of the magnetic field with high permeability material for shielding flow sensor or external magnetic source. 
6, explosion-proof type turbine flow sensor cable connection, must strictly abide by the relevant standards of explosion-proof amplifier. 
Grounding problem 
1, the shielding wire can only be grounded at one end, on display instrument grounding. 
2, reliable grounding should be good, explosion-proof type grounding resistance should be less than 10 Ω. 
The use of turbine flow time must pay attention to the points: 
1, requires clean measured medium, low viscosity, corrosion resistance is small, do not contain impurities, in order to reduce the wear of bearing.If the liquid to be measured a grudge or containing gas, the flowmeter before venting device.In order to avoid damage of impurity in the fluid into the transmitter devious, and in order to prevent turbine stuck, add filters when necessary. 
2, the installation of the flowmeter should avoid vibration, avoid magnetic field and thermal radiation. 
3, medium density and viscosity changes have an impact on flow value, should be revised if necessary.

Seven, the installation place:
1, the installation site should comply with the environmental temperature at 25 ~ 55 ℃ scope, humidity < 80% RH;
2, well ventilated, avoid sun and rain;
3, avoid in piping vibration or the place of stress;
4, avoid easily affected by the strong thermal radiation and radioactive places;
5, must avoid the outside strong electromagnetic field disturbance to the detection signal.If unavoidable, should be set on the amplifier of the flowmeter (sensor) shield and other measures to eliminate interference;
6, in the places where requires explosion-proof, and isolation amplifier should be adopted;
7, the installation position should consider ease of installation, operation, maintenance.
Eight, the deposit should be paid attention to:
Gas turbine flowmeter arrival should be installed in a timely manner.For a battery-powered LRT -i header, not when using the power pin should be placed in the "OFF" (OFF) position, so as to avoid battery depletion affect the service life of the battery.If you want to deposit, please pay attention to the following items:
A) the possible conditions, do not open the packing cases.
B) if already opened the package, or have used the instrument, please put the LRT -i header power jumper wire inserted in the "OFF" position, and use the original packaging.
Location should meet the following conditions:
A) rain moistureproof.
B) small mechanical vibration, avoid collision impact.
C) the temperature in - 30 ~ + 60 ℃.The ideal temperature at 25 ℃ or so.
D) such as stored in outdoor, instrument performance will be affected.So if the instrument to the installation location, will install as soon as possible.
Nine, installation requirements:
1, before installation, pipeline purging, and in case of residual iron affect the normal operation of flowmeter.
2, before the installation, the use of tiny air blows the turbine, the turbine can flexible rotation, no random noise, count rotation is normal, uninterrupted card lag phenomenon, the flowmeter can be used to install.
3, in the middle of the flange and pipe flanges to add sealing washer.
4, before the gas meter shall be equipped with filters, temperament is dirtier occasions should add oil filter, before you place an order, can order to me at the same time, it is strictly prohibited to filter and flowmeter directly connected.
5, gas turbine flowmeter in the installation shall be added as of before and after the valve.
6, flange joint pipe Canon there should be no protrusions.
7, when installation, it is strictly prohibited in the import and export directly to where the flange welding, so as not to burn out flowmeter internal parts.
8, shall be installed in easy maintenance, strong electromagnetic interference and mechanical vibration and thermal radiation effects.
9, unfavorable use frequent interruptions in the flow, and have strong pulsation flow or pressure pulsation of the occasion.
10 and outdoor installation, should be the upper cover, to prevent rain water immersion and strong sunlight affect the service life of the meter.
11, can be horizontal or vertical installation, the fluid flow direction should be consistent with the direction of the logo on the shell, should guarantee the upstream of the flow meter is not less than 10 dn straight pipe, after the table is not less than 5 dn straight pipe.
12, in order not to affect the normal fluid transportation, it is suggested that according to the figure 2 by-pass pipe installation, must close the by-pass pipe valve when normal use.

Ten, use notice:
1, in the low pressure table z pressure leak detection: 1.0 Mpa, or less temperature pressure compensation type leakage pressure no more than three times the upper limit value of the pressure sensor.High pressure casing according to 1.2 times the rated flow pressure leak detection.
2, after installation leak test and installation shall comply with the following provisions:
A close vent valve and the inlet valve first.
B slightly open vent valve, and then slowly open the inlet valve, make the table start slowly, then slowly open the vent valve, make the reach normal operation state, to prevent sudden start, damage to the core.
C at the time of stopping air, need to close the vent valve, then close the inlet valve, each time you start all should abide by the rules.
After 3, install and use, not to touch the bolt, screw, nut fasteners, such as to avoid danger such as leakage, damage.
4, flow meter movement (temperament clean) should be once every six months for cleaning maintenance;(temperament is dirtier) should be cleaning maintenance once a month.
5, if the movement after breakdown maintenance, to be put into use, the need to press zui pressure seal test first.And note within the movement precision instrument transformer oil or oil.

Eleven, measurement range and working pressure:

1, ZW - LWGQ series gas turbine meter:

Nominal diameter
(mm)

model

The standard range
(m³/h)

Extended range
(m³/h)

Stress levels
(Mpa)

High pressure rating
(Mpa)

installation

DN20

LWGQ-20

2.2-25

4-40

1.6

2.5, 4.0

The flange

DN25

LWGQ-25

2.2-25

4-40

1.6

2.5, 4.0

Flange (thread)

DN32

LWGQ-32

5-50

6-60

1.6

2.5, 4.0

The flange

DN40

LWGQ-40

5-50

6-60

1.6

2.5, 4.0

Flange (thread)

DN50

LWGQ-50

6-65

5-70.

1.6

2.5, 4.0

The flange

LWGQ-50

10-100.

8-100.

DN65

LWGQ-65

15-200.

10-200.

1.6

2.5, 4.0

The flange

DN80

LWGQ-80

13-250.

10-160.

1.6

2.5, 4.0

The flange

LWGQ-80

20-400.

The flange

DN100

LWGQ-100

20-400.

13-250.

1.6

2.5

The flange

LWGQ-100

32-650.

The flange

DN125

LWGQ-125

25-700.

20-800.

1.6

2.5

The flange

DN150

LWGQ-150

32-650.

80-1600.

1.6

2.5

The flange

LWGQ-150

50-1000.

The flange

DN200

LWGQ-200

80-1600.

50-1000.

1.6

...

The flange

LWGQ-200

130-2500.

The flange

DN250

LWGQ-250

130-2500.

80-1600.

1.6

...

The flange

LWGQ-250

200-4000.

The flange

DN300

LWQ-300

200-4000.

130-2500.

1.6

...

The flange

320-6500.

1.6

2Turbine flowmeter, ZW - LWGY series liquid:

Meter diameter(mm)

Normal flow range(m3/h)

Expand the flow range(m3/h)

Conventional resistance pressure(MPa)

Special stress levels(MPa)

(Flange connection way)

DN 4

0.04~0.25

0.04~0.4

6.3

12、16、25

DN 6

0.1~0.6

0.06~0.6

6.3

12、16、25

DN 10

0.2~1.2

0.15~1.5

6.3

12、16、25

DN 15

0.6~6

0.4~8

6.3、2.5 (The flange)

4.0、6.3、12、16、25

DN 20

0.8~8

0.45~9

6.3、2.5 (The flange)

4.0、6.3、12、16、25

DN 25

1~10

0.5~10

6.3、2.5 (The flange)

4.0、6.3、12、16、25

DN 32

1.5~15

0.8~15

6.3、2.5 (The flange)

4.0、6.3、12、16、25

DN 40

2~20

1~20

6.3、2.5 (The flange)

4.0、6.3、12、16、25

DN 50

4~40

2~40

2.5

4.0、6.3、12、16、25

DN 65

7~70

4~70

2.5

4.0、6.3、12、16、25

DN 80

10~100

5~100

2.5

4.0、6.3、12、16、25

DN 100

20~200

10~200

2.5

4.0、6.3、12、16、25

DN 125

25~250

13~250

1.6

2.5、4.0

DN 150

30~300

15~300

1.6

2.5、4.0

DN 200

80~800

40~800

1.6

2.5、4.0

Twelve, selection of coding:

1, LWQ series gas turbine flowmeter selection:

type

ZW□□□

-

-

-

-

Instructions (in the wei and control)

LWQ-N

Basic three-wire system output pulse signal,+24VThe power supply

LWQ-A

Current output type4~20mAThe output,+24VThe power supply

LWQ-B

The display type3.6VBattery power

LWQ-C

The scene shows with4~20mAOr pulse output;+24VThe power supply

LWQ-D

Flow transmitter4~20mAThe output,+24VThe power supply.Temperature pressure compensation type

Nominal diameter

20

The normal flow

The scope ofm3/h

2.5-25 m3/h

Expand the flow

The scope ofm3/h

4-40 m3/h

25

2.5-25 m3/h

4-40 m3/h

32

5-50 m3/h

6-60 m3/h

40

5-50 m3/h

6-60 m3/h

50

6-65 m3/h

5-70 m3/h

65

15-200m3/h

10-200m3/h

80

13-250 m3/h

10-160 m3/h

100

20-400 m3/h

13-250 m3/h

125

25-700m3/h

20-800 m3/h

150

32-650 m3/h

50-1000 m3/h

200

80-1600 m3/h

800-1600 m3/h

250

130-2500 m3/h

80-1600 m3/h

300

200-4000 m3/h

320-6500 m3/h

explosion-proof

No mark for the explosion proof type

B

Explosion-proof type

Precision grade

A

precision1.0level

B

precision1.5level

C

precision2.5level

The turbine type

A

Normal flow range

B

Expand the flow range

2, LWGY series liquid turbine flowmeter selection:

type No.

said Ming wei tt&c) in (

ZW-LWGY

-

-

-

-

-

-

-

The male

According to

tong

diameter

4

4mm, the standard range0.04~0.25m3/h, wide range0.04~0.4m3/h

6

6mm, the standard range0.1~0.6m3/hAnd quantity of wheel is wide0.06~0.6m3/h

10

10mm, the standard range0.2~1.2m3/h, wide range0.15~1.5m3/h

15

15mm, the standard range0.6~6m3/hWide range of0.4~8m3/h

20

20mm, the standard range0.8~8m3/h, wide range0.4~8m3/h

25

25mm, the standard range1~10m3/h, wide range0.5~10m3/h

32

32mm, the standard range1.5~15m3/h, wide range0.8~15m3/h

40

40mm, the standard range2~20m3/h, wide range1~20m3/h

50

50mm, the standard range4~40m3/h, wide range2~40m3/h

65

65mm, the standard range7~70m3/h, wide range4~70m3/h

80

80mm, the standard range10~100m3/h, wide range5~100m3/h

100

100mm, the standard range20~200m3/h, wide range10~200m3/h

125

125mm, the standard range25~250m3/h, wide range13~250m3/h

150

150mm, the standard range30~300m3/h, wide range15~300m3/h

200

200mm, the standard range80~800m3/h, wide range40~800m3/h

type

N

The sensor type:+12Vor24VPower supply, signal pulse output three wire system

A

Transmitter type:+24VThe power supply, and two wire output4~20mA

B

Intelligent: lithium-ion batteries, the scene show no signal output

C

Intelligent:+24VPower supply, the scene display and output two wire4~20mA

C1

Intelligent:+24VDisplay with a power supply, the sceneRS486Communication protocol

C2

Intelligent:+24VDisplay with a power supply, the sceneHARTCommunication protocol

Precision grade

05

precision0.5level

10

precision1.0level

vortex wheel

class type

W

Wide-range turbine

S

Standard turbine

material mass

S

304Stainless steel

L

316(L)Stainless steel

explosion-proof

E

Explosion-proof type (Explosion-proof level:ExdⅡBT6)

Classes.

N

conventional

H

High pressure

13, the design and order guidelines:
1, you should pay attention to when ordering turbine flow sensor based on fluid nominal diameter (DN), work pressure (Mpa), temperature (℃), the flow range (m3 / h), fluid types and environmental conditions to choose the appropriate specifications.When have explosion-proof requirements must choose explosion-proof type sensor, and strict attention to explosive-proof grade.
2, need to jiangsu wei display instrument of form a complete set, please refer to the corresponding specifications, choose suitable model, or the information provided by our company technical personnel according to your designs for you selection.Indicate the specifications when you need to use the transmission signal cable length.

After-sales service commitment:

1, from the date of signing of the contract, I in the way provided for products of the company offers free maintenance and maintenance service, commitment to lifelong maintenance service;

2, jiangsu wei measurement and control instrument co., LTD. Sales manager will be in regular contact with customers, understand the usage of products, and solve the problems of the customers to use in the process of produce, free of charge to provide technical support;

3 the man-made damage is found, the warranty period, our company is responsible for the maintenance, and the resulting maintenance fee;

4, product quality problems or are not satisfied with product, the user can choose unconditional return, wei tt&c don't charge any fees, appear quality problem, bear the freight back and forth.