Threaded connection of metal tube rotameter

18金属管转子流量计(螺纹式).jpg
  • 18金属管转子流量计(螺纹式).jpg
ZW-LZ series Threaded connection of metal tube rotameter Can be used for complex flow measurement an…

Description:

Productoverview:

ZW-LZ series Threaded connection of metal tube rotameter Can be used for complex flow measurement and process control environment and various medium conditions, suitable for the measurement of liquid and gas. According to the instructions, the applicable situation and the signal output form classification: pointer type, electric remote type, corrosion type, pressure type, jacket type, explosion-proof type. Support for liquid crystal display, the scene shows the cumulative, instantaneous flow, digital communication, site measurement parameters, the electric remote can output 0 ~ 10mA, 4 ~ 20mA standard analog signal, especially suitable for small diameter and low flow velocity medium measurement.

Two, work principle:

ZW-LZ series threaded metal tube rotameter is the medium to be measured through a conical measuring tube float from upper and lower ends of the differential pressure generated form the upward force, when the float was immersed in a fluid on the lift is greater than the weight of the float, the float will rise, annulus area increases, the annulus fluid velocity decreased immediately float the lower end of the differential pressure is reduced, to float upward force also decreases with, until the lift is equal to the weight of the float when immersed in a fluid, the float will stabilize at a certain height. The float position that corresponds to the size of the measured medium flow. A built-in magnet, the float with the medium moves up and down, with the movement of the float magnetic field changes.

A, for in situ, by rotary magnetic float and the local indicator in the inner magnetic coupling, rotate, and drives the pointer, through the dial indicates the size of the flow.

B, the intelligent, by a solid state magnetic sensor intelligent indicator in the magnetic field will change into electrical signal output by A/D transform, D/A microprocessor, LCD liquid crystal display to show the size and cumulative flow flow.

Three, application:

ZW-LZ series threaded metal tube rotameter has the advantages of simple structure, reliable operation, high accuracy, wide application range. Compared with glass rotor flowmeter can resist high pressure. With local indication, electric remote meter, limit switch alarm, corrosion resistance, damping and explosion-proof type jacket, etc.. Widely used in national defense, chemical, petroleum, metallurgy, electric power, environmental protection, medicine and other departments of the light liquid, gas flow measurement and automatic control.

Four, structure:

By 1, all metal structure, common type, high temperature, high pressure, damping type type type.

2, according to the customer the pipe, the measuring tube structure can be divided into: in the cross, lower upper and lower transverse into cross, right into the left, the left into the right.

3, this series of metal rotameter is mainly composed of two parts: measuring tube and indicator. The measuring tube comprises a conical tube or plate, guide, stopper, float and other components, including magnetic indicator servo system, a pointer, a dial, lines and other components.

Five, main features:

1, all metal structure design, strong, simple, high pressure resistance, high temperature resistance, corrosion resistance, long service life;

2, a short trip, the total height of 250mm, easy to install;

3, mechanical pointer instructions LCD liquid crystal display instantaneous flow, instantaneous flow, cumulative flow;

4, intelligent indicator, using foreign Zui new signal processing chip, modular design, no hysteresis;

5, data recovery, data backup, power down protection and fault self diagnosis function;

6, HART type indicator, two-wire 24VDC power supply, 4-20mA standard current signal output of HART protocol stack;

7, through the HART communication manipulator configuration and parameters can be adjusted;

8, with the PC serial connection, through the Windows program, to achieve on-site calibration of the instrument;

9, can be used in flammable and explosive occasions;

10, vertical and horizontal, upper lower, bottom side, side into the side such as the installation of a variety of forms, flange or threaded connection.

Six, technical parameters:

1, measurement range: water 2.5 ~ 100000l/h (20 C)

2, range: 10:1

3, 1.0, 1.5, precision grade: 2.5

4, working pressure: DN15 DN25, DN50, PN4.0MPazui, DN80, DN100 10.0MPa PN1.6MPazui 6.4MPa

5, medium temperature: -40 to 300 DEG C

6, medium viscosity: DN15: <5mPa - s eta (F15.1 ~ F15.3)

<30mPa - S (F15.4 - F15.8)

DN25: <250mPa - s.

DN50: <300mPa - S - DN150.

7, the ambient temperature: LCD type: -40 to 85 DEG C

Pointer type: -40 to 120 DEG C

8, connection: flange (executive standard DIN2501 or as required by standard flange manufacturing)

9 meter height: 250mm

10, cable interface: M20*1.5

11, power supply: 24VDC two-wire 4 ~ 20mA or 85 ~ 265VAC 50/60Hz (remote type)

12, alarm output: alarm relay output lower or upper limit of instantaneous flow (contact capacity Zui 5A-250VAC) or open collector output (internal impedance Zui 100mA-30VDC 100 ohm).

13, the cumulative output pulse: pulse output, Zui interval of every 10 seconds a pulse (AC) or a pulse every 50 milliseconds.

14, double liquid crystal display: LCD display, display the instantaneous flow and cumulative flow.

15: Exia II CT4, explosion-proof.

16, the measuring tube material: 316 stainless steel (common type) or Teflon liner (corrosion type).

Troubleshooting seven fault:

1, check the alarm set value size SU.

2, check the deviation set D value can not be too large.

3, FUN function, logic function is correct. HA-A said the upper limit is logical. LA-A said the lower limit is.

4, circuit board failure, replace the board.

5, if the lcd barcode indicates the correct output, no action, the cathode can check the external power and external power supply is connected with the anode of instrument power supply.

If the metal tube rotameter jitter in use, according to the jitter amplitude to determine what kind of amplitude jitter.Divided into mild and moderate jitter jitter jitter and jitter.

1, a slight jitter due to fluctuations caused by the media. Can be used to increase the damping of the way to overcome.

2: general, moderate jitter caused by medium flow state. For general gas as the medium pressure instability caused by.The regulator or stabilizer to overcome or increase the damping of gas meter.

3, severe jitter: mainly due to medium gas pulsation, operation pressure instability or give users pressure, temperature, flow meter and the actual state does not have large differences over range instrument.

Metal tube rotameter as long as you use the correct method, usually pay more attention to maintenance, to ensure the normal use of it.

Eighth, the type of flow meter:

1, the metal tube flowmeter will 5 types of products according to the fluid into the entrance and exit direction: vertical type, horizontal type (including the level T, level of spring type), into the side of the bottom and side into the side, side in the type, and each type and standard type, high temperature type, jacket type etc..

2, the level of the installation type meter when the diameter is less than DN50 (inclusive), using T structure, when more than DN50, the spring structure.

3, specifications: standard type flowmeter is connected with GB/T9119-2000 standard flange, flange standard can be made according to customer demand; also can be used threaded connection, clamp connections.

Magnetic filter:

If there is a ferromagnetic particle medium, magnetic filter installed in the entrance before the flowmeter.

Nine, conversion indicator:

The converter is actually the height conversion taper pipe float into the corresponding volume scale. From the output signal: local display and remote signal output type:

1, local display type: by in situ indicator in the follow-up magnetic float and the inner magnetic coupling, and rotate, and the electric pointer by dial indicating the flow

2, intelligent remote type, by the intelligent indicator in the follow-up magnetic float and the inner magnetic coupling, rotate, and drives the magnet and the sensing pointer through a magnetic sensor magnetic field change into electrical signals by A/D conversion, digital filter, D/ microprocessor, A output, LCD liquid crystal display, to show the instantaneous flow and cumulative flow rate.

Ten, product maintenance:

1, in the process of long-term use, the pipeline inevitably there will be ferromagnetic impurity adsorption on the float, it will float stuck or affect the measurement accuracy, from time to time to clean the sensor float; if the installation of magnetic filters should be cleaned regularly.

2, the instrument indicator with electronic components, instrument wiring or demolition of housing, to housing screws, must ensure sealing, prevent impurities, or other material into the water, and also to ensure reliable earthing of the meter casing.

3, the instrument is installed, in use, should pay attention to:

When the valve is opened to avoid the formation of pressure shock makes the float slamming limiting device, causing damage to the instrument, slowly open the valve!

The measurement of gas meter is equipped with gas damper, Z to decrease the float concussion; to ensure the stability of the float, a throttle valve can be installed in the instrument export.

4, for the remote type instrument, we must first ensure that the instrument wiring correctly, keeps power; for dangerous situations, must choose the explosion-proof, and the installation and use of proof required by the.

Eleven, the installation should follow the following steps:

1, before installation to process piping, should remove all installed and checked for any transport damage;

2, in order to prevent in transit, with a rubber band fixed pointer indicator, indicator should open the left cover it gently removed before connection, do not change the pointer position;

3, open the indicator of the right cover, cable through the cable introducing device, according to the wiring diagram wiring rotation transmitter, pressure screw nut, the cable is fixed, and then tighten the fastening screws around the cover, the indicator of waterproof and dustproof;

The upstream and downstream pipeline 4, instrument and instrument should be the same caliber, flange or threaded connection should be matched with the flange and screw instrument, instrument upstream straight pipe length should ensure that the instrument is 5 times the nominal diameter, downstream straight pipe length not less than H2 250mm;

5, due to the use of the instrument measuring mechanism is magnetic transfer, so in order to ensure the performance of rotameter installed around the sensor, at least 100mm, the existence of ferromagnetic material is not allowed;

6 before installation, instrumentation, piping should be purged, prevent the retention of ferromagnetic material in the pipeline is attached to the instrument, the instrument performance, or even damage the instrument. If unavoidable, should be installed at the entrance of the magnetic filter instrument;

7, the measurement of gas meter is calibrated in special pressure, if the gas meter in the outlet directly discharged into the atmosphere, will produce a pressure drop in the float, and lead to data distortion. If this is the working conditions, should install a valve in the instrument exports, so that the required flow value is set, when the float to maintain calibration pressure, gas expansion in valve;

The installation form 8, instrument installation for vertical and horizontal installation, if it is installed in a vertical straight form, should ensure that the instrument is better than 1%; if the level of installation form should ensure that the level of instrument is better than 1%;

9, installed in the pipeline in the instrument should not be subject to stresses, the instrument should be a proper entrance and exit pipe support, can make the instrument in z stress state;

10, product and computer related to calibration and other operations, must be in a safe place.

Twelve, matters needing attention:

1, in the installation before the first pipe flowmeter, dirt, impurities such as welding blowing clean, if the liquid contained solid impurities, should the corresponding filter is installed between the valve and the straight pipe.

2, shielding and grounding system: in order to ensure the ideal electromagnetic compatibility, shielding and grounding is very important, especially the bus cable, cable as far as possible to make continuous shielding.

3, the safety zone, according to the requirements of conventional cable grounding as much as possible.

4, in dangerous places, to ensure that the whole range of bus safe places and dangerous enough, all the potential, in this case, multiple ground is very superior.

Thirteen, the measuring range and the pressure loss:

Decay path

Common path

measuring range

Zui high pressure loss

DN (mm)

DN (mm)

The air (m3/h)

Water (L/h)

The air (KPa)

Water (KPa)

Fifteen

Fifteen

0.05-0.5

1.6-16

Seven

Six point four

0.07-0.7

2.5-25

Seven point one

Six point five

0.11-1.1

4.0-40

Seven point two

Six point five

0.18-1.8

6.0-60

Seven point three

Six point six

0.28-2.8

10-100

Seven point five

Six point six

0.40-4.0

16-160

Eight

Six point eight

0.70-7.0

25-250

Ten point eight

Seven point two

1.00-10

40-400

Ten

Eight point six

1.60-16

60-600

Fourteen

Eleven point one

Twenty-five

-

80-800

-

Fourteen

Twenty-five

3.00-30

100-1000

Seven point seven

Seven

4.50-45

160-1600

Eight point eight

Eight

7.00-70

250-2500

Twelve

Ten point eight

11-110

400-4000

Nineteen

Fifteen point eight

Fifty

-

500-5000

-

Sixteen point two

Fifty

18-180

600-6000

Eight point six

Eight point one

25-250

1000-10000

Ten point four

Eleven

Eighty

40-400

1600-16000

Twelve point six

Seventeen

Eighty

50-500

2000-20000

Fifteen point five

Six point three

One hundred

75-750

2500-25000

Seventeen Point Two

Eight point one

One hundred

100-1000

4000-40000

-

Nine point five

One hundred and twenty-five

160-1600

6000-60000

-

Ten

One hundred and twenty-five

250-2500

8000-80000

-

Sixteen

One hundred and fifty

-

10000-100000

-

Twenty


300-3000

12-120m3/h

-

Twenty


400-4000

15-150m3/h

-

Twenty-five

Fourteen, the selection of encoding:

Specification code

Encoding instructions

The name of the factory

ZW

Zhongwei control

Instrument types

LZ

Metal tube rotameter

structural style

/

In the next

H1

The left into the right

H2

Right into the left out

AA

The side into the side

LA

The bottom side

S

Threaded connection

M

Clamp connection (sanitary polishing tube)

Size code

- XXX

Cases: Fifteen Express DN15

The measuring tube material

R1

1Cr18Ni9Ti , Three hundred and four

R0

0cr18Ni12M02Ti , Three hundred and sixteen

RL

316L

RP

(PTFE F-lined )

Ni

(Hasloy Hastelloy C)

Ti

(TitHXium titanium alloy )

Media type

L

liquid

G

Gas

  working temperature 

H1

- 40 C -100 C

H2

- 80 C -250 C

Rated pressure

One point six

1.6MPa

Four

4.0MPa

Sixteen

16MPa

Twenty-five

25MPa

Additional structure

O

nothing

Z

Damping type

T

Jacket

H

High temperature type

Y

High pressure type

Form indicator

M1

Local indication

M2

Intelligent

M3

Intelligent ( Flameproof )

The output signal in the communication

N

nothing

A

4-20mA

H

4-20mA+Hart

Alarm controller

K1

The upper limit alarm

K2

Alarm

K12

The upper and lower limit alarm

Explosion proof type

E1

Intrinsically safe EX IaCT5 Square shell

E2

Flameproof EXiibt4 Park Housing

Power supply

Zero

24VDC (can be omitted )

Two

220VAC

Three

Three point six Volt lithium battery

Fifteen, design selection and ordering information:

1, please provide the pipe size, wall thickness, material requirements and accuracy requirements;

2, please provide the liquid or gas, such as the name, commonly used commonly used temperature, working pressure, common flow, density, viscosity etc.;

3, corrosion proof, should indicate the measured corrosion and concentration etc.;

4, if there are special to show, please contact our sales department.

Customer service service commitment:

1, our series of metal tube flowmeter series product quality problems within twelve months, responsible for free maintenance.

2, the warranty period if the products of our company have any quality problems, our company is responsible for free repair or replacement.

3, the warranty period if the user use undeserved, cause damage to the product, our company is responsible for the maintenance, collect the damaged parts cost.

4, after the warranty period of product quality problems, our company is responsible for the maintenance, collect the damaged parts cost.

5, if the user needs to our service, our company in the province of the user service personnel arrived at the scene within 12 hours, arrived at the scene 48 hours outside users.