ZW-LZ series sanitary type metal tube rotameter with local display and intelligent remote type, with…
Description:
Productoverview:
ZW-LZ series sanitary type metal tube rotameter with local display and intelligent remote type, with a pointer to display the instant / cumulative flow of liquid crystal display, upper and lower limit alarm output, pulse output, batch control standard, two-wire 4-20mA current output and other forms, for the user to provide a very broad space, in addition to the the instrument adopts the advanced micro processing chip and high quality industrial central components, ensure the flow in all kinds of applications with good performance.
Two, work principle:
ZW-LZ series sanitary type metal tube rotameter is the medium to be measured through a conical measuring tube float from upper and lower ends of the differential pressure generated form the upward force, when the float was immersed in a fluid on the lift is greater than the weight of the float, the float will rise, annulus area increases, the annulus fluid velocity decreased immediately float the lower end of the differential pressure is reduced, to float upward force also decreases with, until the lift is equal to the weight of the float when immersed in a fluid, the float will stabilize at a certain height.The float position that corresponds to the size of the measured medium flow.A built-in magnet, the float with the medium moves up and down, with the movement of the float magnetic field changes.
A, for in situ, by rotary magnetic float and the local indicator in the inner magnetic coupling, rotate, and drives the pointer, through the dial indicates the size of the flow.
B, the intelligent, by a solid state magnetic sensor intelligent indicator in the magnetic field will change into electrical signal output by A/D transform, D/A microprocessor, LCD liquid crystal display to show the size and cumulative flow flow.
Three, application:
ZW-LZ series sanitary type metal tube rotameter is suitable for small flow, low Reynolds number of medium flow measurement is widely used in food, chemical, petroleum, metallurgy, electric power, environmental protection, medicine and light industry and other departments of the liquid flow measurement and automatic control.
Four, main features:
1, suitable for small diameter and low flow velocity medium measurement;
2, reliable operation, low maintenance, long service life;
3, for the straight pipe is not high, wide flow ratio 10:1
4, two large LCD display, optional field instantaneous / total flow rate display, with backlight;
5, single axis sensitive indicator;
6, all metal structure, suitable for high temperature, high pressure and strong corrosive medium;
7, can be used in flammable and explosive occasions.
Five, technical parameters:
1, measuring range: water (20 C) 60~55000L/h see flow meter;
2, than the standard range: 1:10;
3, accuracy: 1.5;
4, pressure rating: DN25~DN100 PN0.6;
5, installation: installed vertically in the bottom medium;
6, pressure loss: 7kPa-70kPa;
7, medium temperature: Standard Type: -80 ~+200 DEG c;
8, medium viscosity: DN25: < 250mPa.s;
DN50~DN100: < 300mPa.s;
9, the ambient temperature: liquid crystal type -30 ~+55 DEG c;
10, the pointer type: -40 ~+120 DEG c;
11, the connection form: fast loading chuck;
12, cable interface: M20 * 1.5;
13, power supply: 24VDC standard two-wire 4~20mA (10.8VDC~36VDC);
14, AC type: 85~265VAC 50HZ;
15, battery type: 3.6V@4AH Ni MH battery;
16, alarm output: the lower or upper limit of instantaneous flow alarm;
17, standard type: open collector output (internal impedance 100mA@30VDC 100 ohm);
18, special type: relay output (contact capacity 5A@250VAC);
19, the cumulative output pulse: pulse output, small interval of 50 ms;
20, the liquid crystal display: instantaneous flow display value range: 0~50000;
21, the cumulative flow display value range: 0~99999999;
22, protection grade: IP65;
23, intrinsically safe explosion proof mark: ia II CT6 flameproof D II BT4.
Six, the type of flow meter:
1, the metal tube flowmeter will 5 types of products according to the fluid into the entrance and exit direction: vertical type, horizontal type (including the level T, level of spring type), into the side of the bottom and side into the side, side in the type, and each type and standard type, high temperature type, jacket type etc..
2, the level of the installation type meter when the diameter is less than DN50 (inclusive), using T structure, when more than DN50, the spring structure.
3, specifications: standard type flowmeter is connected with GB/T9119-2000 standard flange, flange standard can be made according to customer demand; also can be used threaded connection, clamp connections.
Magnetic filter:
If there is a ferromagnetic particle medium, magnetic filter installed in the entrance before the flowmeter.
Seven, conversion indicator:
The converter is actually the height conversion taper pipe float into the corresponding volume scale.From the output signal: local display and remote signal output type:
1, local display type: by in situ indicator in the follow-up magnetic float and the inner magnetic coupling, and rotate, and the electric pointer by dial indicating the flow
2, intelligent remote type, by the intelligent indicator in the follow-up magnetic float and the inner magnetic coupling, rotate, and drives the magnet and the sensing pointer through a magnetic sensor magnetic field change into electrical signals by A/D conversion, digital filter, D/ microprocessor, A output, LCD liquid crystal display, to show the instantaneous flow and cumulative flow rate.
Eight, product maintenance:
1, in the process of long-term use, the pipeline inevitably there will be ferromagnetic impurity adsorption on the float, it will float stuck or affect the measurement accuracy, from time to time to clean the sensor float; if the installation of magnetic filters should be cleaned regularly.
2, the instrument indicator with electronic components, instrument wiring or demolition of housing, to housing screws, must ensure sealing, prevent impurities, or other material into the water, and also to ensure reliable earthing of the meter casing.
3, the instrument is installed, in use, should pay attention to:
When the valve is opened to avoid the formation of pressure shock makes the float slamming limiting device, causing damage to the instrument, slowly open the valve!
The measurement of gas meter is equipped with gas damper, Zui to decrease the float concussion; to ensure the stability of the float, a throttle valve can be installed in the instrument export.
4, for the remote type instrument, we must first ensure that the instrument wiring correctly, keeps power; for dangerous situations, must choose the explosion-proof, and the installation and use of proof required by the.
Nine, matters needing attention:
1, in the installation before the first pipe flowmeter, dirt, impurities such as welding blowing clean, if the liquid contained solid impurities, should the corresponding filter is installed between the valve and the straight pipe.
2, shielding and grounding system: in order to ensure the ideal electromagnetic compatibility, shielding and grounding is very important, especially the bus cable, cable as far as possible to make continuous shielding.
3, the safety zone, according to the requirements of conventional cable grounding as much as possible.
4, in dangerous places, to ensure that the whole range of bus safe places and dangerous enough, all the potential, in this case, multiple ground is very superior.
10, flow parameter table:
Health clamp diameter DN(Mm)
Flow range
Clamp size
Supporting pipe size
water(20CL/h)
Phi(Mm)
Phi(Mm)
Twenty-five
Sixty~Six hundred
Fifty point five
Forty-eight
One hundred~One thousand
One hundred and sixty~One thousand and six hundred
Two hundred and fifty~Two thousand and five hundred
Thirty-two
One hundred and sixty~One thousand and six hundred
Fifty point five
Forty-eight
Two hundred and fifty~Two thousand and five hundred
Four hundred~Four thousand
Five hundred and fifty~Five thousand and five hundred
Forty
Two hundred and fifty~Two thousand and five hundred
Fifty point five
Forty-eight
Four hundred~Four thousand
Five hundred and fifty~Five thousand and five hundred
Fifty
Four hundred~Four thousand
Sixty-four
Fifty-seven
Six hundred~Six thousand
One thousand~Ten thousand
Sixty-five
One thousand~Ten thousand
Ninety-one
Seventy-six
One thousand and six hundred~Sixteen thousand
Two thousand and five hundred~Twenty-five thousand
Eighty
One thousand and six hundred~Sixteen thousand
One hundred and six
Eighty-nine
Two thousand and five hundred~Twenty-five thousand
3500~35000
One hundred
Four thousand~Forty thousand
One hundred and nineteen
One hundred and eight
Five thousand and five hundred~Fifty-five thousand
Eleven, the selection of encoding:
Specification code
Encoding instructions
The name of the factory
ZW
Zhongwei control
Instrument types
LZ
Metal tube rotameter
structural style
/
In the next
H1
The left into the right
H2
Right into the left out
AA
The side into the side
LA
The bottom side
S
Threaded connection
M
Clamp connection (sanitary polishing tube)
Size code
- XXX
Cases:FifteenExpressDN15
The measuring tube material
R1
1Cr18Ni9Ti,Three hundred and four
R0
0cr18Ni12M02Ti,Three hundred and sixteen
RL
316L
RP
(PTFEF-lined)
Ni
(HasloyHastelloyC)
Ti
(TitHXiumtitanium alloy)
Media type
L
liquid
G
Gas
working temperature
H1
- 40C-100C
H2
- 80C-250C
Rated pressure
One point six
1.6MPa
Four
4.0MPa
Sixteen
16MPa
Twenty-five
25MPa
Additional structure
O
nothing
Z
Damping type
T
Jacket
H
High temperature type
Y
High pressure type
Form indicator
M1
Local indication
M2
Intelligent
M3
Intelligent(Flameproof)
The output signal in the communication
N
nothing
A
4-20mA
H
4-20mA+Hart
Alarm controller
K1
The upper limit alarm
K2
Alarm
K12
The upper and lower limit alarm
Explosion proof type
E1
Intrinsically safeEXIaCT5Square shell
E2
FlameproofEXiibt4Park Housing
Power supply
Zero
24VDC(can be omitted)
Two
220VAC
Three
Three point sixVolt lithium battery (battery powered)
Twelve, design selection and ordering information:
1, please provide the pipe size, wall thickness, material requirements and accuracy requirements;
2, please provide the liquid or gas, such as the name, commonly used commonly used temperature, working pressure, common flow, density, viscosity etc.;
3, corrosion proof, should indicate the measured corrosion and concentration etc.;
4, if there are special to show, please contact our sales department.
Customer service service commitment:
1, our series of metal tube flowmeter series product quality problems within twelve months, responsible for free maintenance.
2, the warranty period if the products of our company have any quality problems, our company is responsible for free repair or replacement.
3, the warranty period if the user use undeserved, cause damage to the product, our company is responsible for the maintenance, collect the damaged parts cost.
4, after the warranty period of product quality problems, our company is responsible for the maintenance, collect the damaged parts cost.
5, if the user needs to our service, our company in the province of the user service personnel arrived at the scene within 12 hours, arrived at the scene 48 hours outside users.