Address:No. 4-4 East West Road, Jinhu County Industrial Concentration Zone, Huaian City, Jiangsu Province
Tel:0517-86918678
Fax:0517-86918679
Email:zhongweick@163.com
Post Encoding:211600
The thermal resistance of remote bimetal thermometer
WSSP series The thermal resistance of remote bimetal thermometer And the development of a series of …
Description:
Productoverview: WSSP seriesThe thermal resistance of remote bimetal thermometerAnd the development of a series of temperature instruments, to overcome the traditional thermometer single expansion performance and short service life and poor reliability, the disadvantages of large volume temperature package.The temperature sensor of steam and gas volume respectively than the traditional pressure type thermometer reduced by 30 and 60 times, creatively or platinum copper heat resistance temperature sensor mounted on the measuring element, realize the function of temperature measurement of electromechanical integration.Formed by the liquid pressure type temperature is extremely basic temperature measuring instrument for remote transmission, explosion-proof, anti-corrosion, shockproof, electrical connector, temperature signal transmission and other functions, a series of temperature meter.
Two, work principle: WSSP seriesThe thermal resistance of remote bimetal thermometerIs a kind of thermal resistance temperature sensor and WSSP type thermometer temperature package installed in the same temperature element within a temperature measuring instrument.Two measurement systems are independent of each other, do not affect the temperature measurement accuracy and their.The outstanding advantages of this instrument is just a single point temperature measurement can double indicate the measured temperature, the pressure for the instructions, the thermal resistance signal for transmission to the control room (about 500 meters), with two meter line, to control and adjust the measured temperature, the two temperature measurement system to monitor each other, improve the temperature measurement accuracy and reliability. Three, application: WSSP seriesThe thermal resistance of remote bimetal thermometerNot only meet the temperature requirements, but also to meet the needs of long-distance transmission, remote double metal thermometer can directly measure the production process in a variety of -40 ~ +600 temperature range of liquid, gas and steam medium and solid surface temperature measurement. Four, main features: 1, a temperature probe of small, high sensitivity, linear scale, long service life etc.. 2, with remote output resistance signal (PT100), shock resistance, corrosion resistant function and large power switch signal etc.. 3, structure and international similar products, can replace imported. Five, technical parameters: 1, the temperature range of -40 to 600 DEG c; 2, the ambient temperature of -10 to 55 DEG c; 3, the relative humidity is less than or equal to 95%; 4, precision grade 1.5; 5, the time constant is less than or equal to 60S; 6, the probe voltage is less than or equal to 6Mpa; 7, probe diameter: 8-10mm diameter, movable external thread: M27*2; 8, thermal resistance PT100 (single channel, dual output or 4-20mA output current).
Six, matters needing attention: 1, in storage, installation, use and transport process, to avoid the collision protection tube, prohibit the protection tube bending and deformation. 2, installation, non twist instrument shell. 3, the instrument shall work normally in the -30 to 80 DEG C ambient temperature. 4, the instrument can often work in the temperature scale range 1/2 ~ 3/4. 5, the protective tube immersionmedium measured in length must be greater than the length of the sensing element, this is done in order to ensure the accuracy of measurement. 6, do not use media exposure in measuring container. 7, the use of the process should pay attention to check in time, probably in about 6 months.
Seven, technical data sheet:
typeNumber
Graduation
The measuring range of C
Precision grade
Protection tube material
Insertion length
WSSE-401
E
-80~+40
Forty~+80
Zero~Fifty
Zero~One hundred
Zero~One hundred and fifty
Zero~Two hundred
Zero~Three hundred
Zero~Four hundred
Zero~Five hundred
One point five
1Cr18Ni9Ti
Three hundred and four;
Three hundred and sixteen;
316L;
Harrington
C-276
One hundred and fifty
Two hundred
Three hundred
Four hundred
Five hundred
Seven hundred and fifty
One thousand
WSSE-501
WSSE-411
WSSE-511
WSSE-481
WSSE-581
WSSP-401
Pt100
WSSP-501
WSSP-411
WSSP-511
WSSP-481
WSSP-581
WSSXE-401
E
WSSXP-401
Pt100
WSSXE-411
E
WSSXP-411
Pt100
WSSXE-481
E
WSSXP-481
Pt100
Eight, temperature range:
The measuring range of C
Applicable scope
Industrial and commercial
Laboratory and small
-80 ~ +40
B.
B.
-40 ~ +80
B.
B.
0~50
B.
B.
0~100
B.
B.
0~150
B.
B.
0 ~ 200
B.
B.
0 ~ 300
B.
B.
0~400
B.
B.
0~500
B.
B.
Remarks: OneThe insertion depth of the thermometer,LMust be greater thanLmin,LAlternative series:One hundred,One hundred and fifty,Two hundred,Two hundred and fifty,Three hundred,Four hundred,Five hundred,Seven hundred and fifty,One thousand,One thousand two hundred and fifty,One thousand and five hundred,Two thousandSo, also can be customized according to user needs. TwoAnd the protective tube diameterPhi 6,Phi 8The series of thermometers, the time constant is less thanFortySecond, diameterPhi 10When the time constant is less thanThirtySeconds. ThreeProtect the stainless steel tube1Cr18Ni9TiManufacturing, at room temperature under high working pressure is Zui6.4MPa.
Nine, shape and size:
structural style
A
B
C
E
L
D
The axial type double metal thermometer
Sixty-five
Twenty-three
Seventy-three
-
Seventy-five One hundred One hundred and fifty Two hundred Three hundred Four hundred Five hundred Seven hundred and fifty One thousand
Phi 6 Phi 8 Phi 10
One hundred and five
Twenty-three
Seventy-three
-
One hundred and fifty-five
Twenty-three
Seventy-three
-
The radial type double metal thermometer
Sixty-five
Fifty
One hundred and ten
Thirty-four
One hundred and five
Fifty
One hundred and ten
Thirty-four
One hundred and five
Fifty
One hundred and ten
Thirty-four
135 degrees to type double metal thermometer
One hundred and five
Twenty-three
Eighty-five
-
One hundred and fifty-five
Twenty-three
Eighty-five
-
Universal type double metal thermometer
One hundred and five
Twenty-three
One hundred and seventy-eight
One hundred and twenty
One hundred and fifty-five
Twenty-three
One hundred and seventy-eight
One hundred and twenty
Ten, the selection of encoding:
Serial number
category
Specifications
Explain
WSS
P
Thermal resistance type
Double metal thermometer (Zhongwei control)
Dial diameter
-3
Phi 60
-4
Phi 100
-5
Phi 150
Dial form
Zero
The axial (straight)
One
Radial (angle)
Two
135 degrees to (an obtuse type)
Eight
Universal (angle adjustable type)
Fixed form
Zero
No fixed device
One
Movable external thread
Two
The movable internal thread
Three
Fixed thread
Four
Fixed flange
Five
Card set screw
Six
Sleeve flange
Type of protection
-
Ordinary type
W
Type of protection
F
Anti-corrosion type
Cases:The axial type, PT100 type, I temperature range of 0-400 DEG c..Activity thread M27x2, protective tube insertion length is 316 300mm. WSSP-401 0-400 l=300mm I: C protection tube 316.
The design and selection of notes: 1, model; 2, the thermal resistance index number; 3, the thermal resistance precision grade; 4, double metal thermometer precision grade; 5, temperature range; 6, fixed form; 7, protection tube material; 8, the inserted length. Customer service service commitment: 1, my company sold ZhongweiThermometer seriesThe product within twelve months of quality problems, responsible for free maintenance. 2, the warranty period if the products of our company have any quality problems, our company is responsible for free repair or replacement. 3, the warranty period if the user use undeserved, cause damage to the product, our company is responsible for the maintenance, collect the damaged parts cost. 4, after the warranty period of product quality problems, our company is responsible for the maintenance, collect the damaged parts cost. 5, if the user needs to our service, our company in the province of the user service personnel arrived at the scene within 12 hours, arrived at the scene 48 hours outside users.